Technical Specifications |
Voltage DC |
|
Range: 0 to ±1020V |
|
Best 1 Year Specification: ±0.01% of setting |
|
Current DC |
|
Range: 0 to ±20.5A |
|
Best 1 Year Specification: ±0.05% of setting |
|
Resistance |
|
Range: 0 to 190 MO, 19 values in x1 and x1.9 |
|
Best 1 Year Specification: ±0.025% of setting |
|
Voltage AC |
|
Range: 1 mV to 1020V
45Hz to 1 kHz, Sine |
|
Best 1 Year Specification: ±0.1% of setting |
|
Current AC |
|
Range:29 mA to 20.5A
45 Hz to 1 kHz, Sine |
|
Best 1 Year Specification: ±0.1% of setting |
|
DC Power |
|
Range:10.9 mW to 20.5 kW |
|
Best 1 Year Specification: ±0.11% of setting |
|
AC Power |
|
Range: 10.9 mW to 20.5 kW
45-65 Hz, PF=1 |
|
Best 1 Year Specification: ±0.36% of setting |
|
Phase |
|
Range: 0 to ± 179.99°
45 Hz to 1 kHz |
|
Best 1 Year Specification: ±0.25° of setting |
|
Frequency |
|
Range: 45 Hz to 1 kHz |
|
Best 1 Year Specification: ±50 ppm of setting |
|
General Specifications |
Warm up Time |
|
Twice the time since last warmed up, to a maximum of 30 minutes |
|
Settling Time |
|
Less than 5 seconds for all functions and ranges except as noted |
|
Standard Interfaces |
|
Temperature Performance |
Operating: 0°C to 50°C |
Calibration (tcal):15°C to 35°C |
Storage:-20°C to +70°C |
|
Temperature Coefficient |
|
Temperature coefficient for temperatures outside tcal ±5 °C is 10 %
of the stated specification per °C for temperatures in the range of 0 °C
to 35 °C. Above 35 °C, the temperature coefficient is 20 % of the
stated specification per °C. |
|
Relative Humidity |
Operating: <80% to 30°C, <70% to 40°C, <40% to 50°C |
Storage:<95%, non-condensing |
|
Altitude |
Operating: 2,000m (6,500 ft) maximum |
Non-operating: 12,200m (40,000 ft) maximum |
|
Safety |
Meets EN. 61010-1:2001, CAN/CSA-C22.2 No. 61010-1-04, UL 61010-1: 2004 |
|
Analog Low Isolation |
|
EMC |
|
Line Power |
Line Voltage (selectable): 100V, 120V, 220V, 240V |
Line Frequency: 47 to 63 Hz |
Line Voltage Variation: + 10% about line voltage setting |
|
Power Consumption |
|
Dimensions |
Height:19.3 cm (7.6 in) |
Width:43.2 cm (17 in), 44.3 cm (17.5 in) including handles |
Depth:53.8 cm (21.2 in) |
|
Weight (without options) |
|
Absolute Uncertainty Definition |
The specifications include stability, temperature coefficient,
linearity, line and load regulation, and the traceability of the
external standards used for calibration. It is not necessary to add
anything to determine the total specification for the temperature range
indicated. |
|
Specification Confidence Interval |
|
Accesorios Opcionales:
5500A/COIL (Bobina de Corriente de 50 Vueltas) 5080A/MEG (Opción de calibración de medidor de megahomios 5080A-MEG) 5080A/CASE (Maletín Transportador)
|
|
|
|
|
|
|
|
|
|
|
|
|
Descripcion
Características de los Calibradores 5080A con Circuitos de Protección Contra Voltajes Peligrosos a la Entrada.El calibrador Multiproducto Fluke 5080A calibra sus equipos análogos y digitales de manera exacta y ecónomica. Su capacidad en corriente y voltaje hace la calibracion fácil y precisa. Además cuenta con una circuiteria de entrada que lo protege de voltajes dañinos en las entradas.
• Calibra una amplia variedad de equipos de pruebas eléctricos • Los robustos circuitos de protección evitan daños costosos por errores del operario • Las asas para transporte rediseñadas facilitan el transporte del 5522A • Maletín de transporte resistente con ruedas y asas incorporadas y puertas de acceso delanteras/traseras extraíbles para calibrar in situ en cualquier entorno • Sorprendentemente accesible
El calibrador 5080A abarca muchas de las herramientas para pruebas electrónicas que utiliza para mantener su compañía funcionando, incluyendo: • Medidores portátiles y de banco (analógicos y digitales) de hasta 4½ dígitos • Sondas de corriente y medidores de sonda • Termómetros RTD y termopar • Calibradores de procesos • Registradores de datos • Registradores de banda y de gráficos • Vatímetros • Analizadores de armónicos de alimentación • Medidores de panel • Multímetros gráficos • Analizadores de calidad de alimentación (con opción) • Osciloscopios portátiles y de banco analógicos o digitales hasta 200 MHz (con opciones) • …y más, incluyendo transductores y manómetros, y medidores de potencia trifásica
|
|
|
|
|
|
|
|
|