DC voltage, output | Range [1] | Absolute Uncertainty, ± (ppm of output + μV), 1-yr | Resolution | 0 to 100 mV | 30 | 3 | 1 μV | 0 to 1 V | 30 | 10 | 10 μV | 0 to 10 V | 30 | 100 | 100 μV | 0 to 100 V | 30 | 1 mV | 1 mV | TC Output and Input | –10 mV to 75 mV | 30 | 2 | 10 Ω |
[1] All outputs are positive only unless otherwise noted DC voltage, isolated input | Range | Absolute Uncertainty, ± (ppm of reading + mV), 1-yr | Resolution | 0 to 10 V | 50 | 0.2 | 100 μV | 10 V to 100 V | 50 | 2.0 | 1 mV |
DC Current, output | Range[1] | Absolute Uncertainty, ± (ppm of reading + μA), 1-yr | Resolution | 0 to 100 mA | 50 | 1 | 1 μA |
[1] For line voltages less than 95 V, ±100 ppm of reading
DC current, isolated input | Range | Absolute Uncertainty, ± (ppm of reading + μA), 1-yr | Resolution | 0 mA to 50 mA | 100 | 1 | 0.1 μA | 0 mA to 24 mA [1][2] (Loop Power) | 100 | 1 | 0.1 μA |
[1] Loop Power: 24 V ±10% [2] HART Resistor: 250 Ω ± 3%
Resistance, output | Range | Absolute Uncertainty, tcal ± 5 °C, ± Ohms, 1-yr | Resolution | Nominal current | 5 Ω to 400 Ω | 0.015 | 0.001 Ω | 1 to 3 mA | 5 Ω to 4 kΩ | 0.3 | 0.01 Ω | 100 μA to 1 mA |
Resistance, input | Range | Absolute Uncertainty, ± (ppm of reading + Ω), 1-yr | Resolution | 0 Ω to 400 Ω | 20 | 0.004 | 0.001 Ω | 0 Ω to 4 kΩ | 20 | 0.04 | 0.01 Ω |
Sample thermocouple accuracy, input/output (does not include all available TC types)[1] | TC type | Temperature Range (°C) | Absolute Uncertainty, tcal ±5 °C, ± (°C), 1-yr [2] | Min | Max | J | -210 | 1200 | 0.09 | K | -250 | 1372 | 0.1 | S | -50 | 1767 | 0.29 | T | -250 | 400 | 0.11 |
[1] See extended specifications for all TC types (B,C,E,J,K,L,N,R,S,T,U,XK,BP) [2] Best accuracy within specified TC temperature range
Sample RTD and thermistor, output (does not include all available RTD types)[1] | RTD Type | Temperature Range (°C) | Absolute Uncertainty, tcal ±5 °C, ± (°C), 1-yr | Min | Max | Pt 385, 100 Ω | -200 | 630 | 0.05 | YSI 400 | 15 | 50 | 0.007 |
[1] See extended specifications for all RTD types: Pt-100 (385, 3926, 3916), Pt- 200, Pt-500, Pt-1000, Ni-120, Cu-427, SPRT
Sample RTD and thermistor, intput (does not include all available RTD types)[1] | RTD Type | Temperature Range (°C) | Absolute Uncertainty, tcal ±5 °C, ± (°C), 1-yr | Min | Max | Pt 385, 100 Ω | –80 | 100 | 0.020 | 100 | 300 | 0.024 | YSI 400 | 15 | 50 | 0.007 |
[1] See extended specifications for all RTD types: Pt-100 (385, 3926, 3916), Pt- 200, Pt-500, Pt-1000, Ni-120, Cu-427, SPRT
General specifications | Standard interface | RS-232, IEEE-488 (GPIB) | Temperature performance | Operating: | 0 °C to 50 °C | Calibration (tcal): | 18 °C to 28 °C | Storage: | –20 °C to 70 °C | Electromagnetic compatibility | CE: Conforms to EN61326; operation in controlled EM environments | Temperature coefficient | Temperature coefficient for temperatures outside tcal 5 °C is 10% of the 90-day specification (or 1 year if applicable) per °C | Relative humidity | Operating: | <80% to 30 °C | <70% to 40 °C | <40% to 50 °C | Altitude | Operating: | 3,000 m (9,800 ft) max | Non-operating: | 12,200 m (40,000 ft) max | Safety | EN/IEC 61010-1:2010 3rd Edition, UL 61010-1:2012, CAN/CSA 22.2 No. 61010-1-12 | Analog low isolation | 20 V | Line power | 120 V~: | 100 V to 120 V | 240 V~: | 220 V to 240 V | Line frequency | 47 Hz to 63 Hz | Line voltage variation | ± 10 % about setting | Power consumption | 15 VA maximum | Dimensions | Height: | 14.6 cm (5.75 in) | Width: | 44.5 cm (17.5 in) | Depth: | 29.8 cm (11.75 in) | Weight (without options) | 4.24 kg (9.35 lb) |
|
|
|
|
|
|
|
|
|
|
|
|
|
Descripcion
Versátil, preciso y de calidad, características combinadas en una sola herramienta de banco para el proceso de calibración
El calibradr de FLuke 7526A calibrador de procesos de precision ofrece el mejor balance entre enconomia y precision para calibracion de temperatrua y procesos de presion de instrumentacion. Facilmente calibra las lecturas hechas por RTD y termocuplas, galgas de presion, transmisores de temperatura, simuladores de procesos digitales, data loggers, multimetros y mas. - Mide y provee, voltaje de DC, corriente, resistencia, RTDs y termocuplas
- Mediciones precisas de presion usando el Fluke 700 o los modulos de presion de la serie 525A-P
- Incluye una fuente de voltaje de DC de 24 v
- Mide un lazo de corriente de 4-20mA
- Incluye una funcion automatizada de cambio de funcion
- Acepta coeficientes ITS-90 para mediciones exactas de SPRT
- Compatible con MET/CAL® Calibration Software
|
|
|
|
|
|
|
|
|