Rated current: | 500 A AC/DC | Max. allowable input: | 1000 A DC (requires derating) | Frequency characteristics: | Amplitude: DC to 100 kHz Phase: DC to 100 kHz | Amplitude and Phase accuracy: | DC ±0.3 % rdg. ±0.02 % f.s. (Phase: Not defined) DC < f ≤ 100 Hz ±0.3 % rdg. ±0.01 % f.s. (Phase: ±0.1 deg) Defined to 100 kHz | Power consumption: | 7 VA max. (at 500 A/55 Hz, ±12 V power requirement) | Output voltage: | 4 mV/A (voltage output with the Sensor Unit 9555-10, use with a device having a 1 MΩ input resistance or higher) | Core diameter: | φ 50 mm (1.79 in) | Operating temperature, humidity: | -40 °C to +85 °C (-40 °F to 185 °F), 80% rh or less (with no condensation) | Power supply: | DC ±11 V to ±15 V (Power suppled via the 9555-10, which supports 100 to 240 V AC) | Dimensions and mass: | 238 mm (9.37 in)W × 116 mm (4.57 in)H × 35 mm (1.38 in)D, cord length: 3 m (9.84 ft), 870 g (30.7 oz) | Compatible models | CT6845 Combination status | CT6845-05 Combination status | Model 3390 | OK | △ (Requires Conversion Cable CT9901) | Model PW6001 | △ (Requires Conversion Cable CT9900) | OK | Input unit 9602 for 3193-10/ 3193/ 3194 | OK (In combination, recognized as 500A AC/DC) | △ (In combination, recognized as 500A AC/DC, requires Conversion Cable CT9901) | Model 8971 for the MR8847-01s | △ (In combination, recognized as 9279 [500 A rated], requires Conversion Cable 9318) Maximum number of connectable sensors when using MR8827 or MR8847-51/52/53: 5 | △ (In combination, recognized as 9279 [500 A rated], requires Conversion Cable 9318 and CT9901) Maximum number of connectable sensors when using MR8827 or MR8847-51/52/53: 5 | Model 8940 for the Memory Hicorders | △ (In combination, recognized as 500A AC/DC, requires Conversion Cable 9705 and 9318) | △ (Requires Conversion Cable 9705, 9318 and CT9901)
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Descripcion
- Ideal para uso en pruebas ambientales con un amplio rango de temperatura de -40°C a 85°C
- Alta precisión con un diseño tipo gancho, precisión de amplitud de ±0.3%, precisión de fase de ±0.1°
- Amplio ancho de banda DC a 100 kHz
- Diámetro grande para cables gruesos y doble cableado
- Reducción de effectos por campos magnéticos, posición de conductor, y ruido de cables cercanos
- Para medición de eficiencia de carga y descarga en baterías EV/HEV y evaluación de eficiencia de conversión de inversor y acondicionador de potencia
|
|
|
|
|
|
|
|
|