GX2002
The GX2002 PXI RF Power Analyzer module performs multiple
|
The GX2002 consists of two inputs, an SMA signal input and an SMB trigger input/output. The trigger input is TTL compatible and can be programmed to be positive or negative edge sensitive. Alternatively, triggering of a measurement or acquisition can be based on a trigger created by the RF input signal. A flexible data acquisition system allows the user to program the aperture acquisition window, the aperture delay, as well as the signal threshold level, providing flexibility to acquire and analyze RF bursts from 100 ns to 1.25 seconds in length. The module’s on-board DSP provides that ability to perform all measurements including the input signal's pulse-width, PRI, frequency, and peak power in less than 100 microseconds.
The GX2002 is supplied with a virtual instrument panel, which includes 32-bit DLL driver libraries and documentation. The virtual panel can be used to interactively program and control the instrument from a window that displays the instrument’s current settings and measurements.
In addition, various interface files provide access to the library for programming tools and languages such as ATEasy, Microsoft® and Borland® C/C++, Microsoft Visual Basic®, LabVIEW, Borland Delphi, and more.
| ATE Systems | |
| Radar test systems | |
| Mobile and base radio test systems | |
| WLAN applications | |
| RF semiconductor test applications |
| Input Characteristics | |
| Impedance | 50 Ù VSWR: < 1.2:1, 100 MHz to 2.5 GHz VSWR: < 1.4:1, 2.5 GHz to 3 GHz |
| Coupling | AC |
| Ext. Trigger | Positive or Negative edge, selectable Ext Input level: TTL compatible, +/- 5 volts max PXIbus TRG0-7 PXIbus STAR TRG |
| Signal Range | -34 dBm to +16 dBm |
| Internal Trigger (Threshold) Level | |
| Range | -39 dBm to +20 dBm |
| Auto Trigger | Automatic threshold detection, level is automatically set approximately 6 dB below detected peak power |
| External Trigger Output | |
| Output Signal Characteristics | Ext output: DC Coupled, TTL compatible |
| Measurement Functions | |
| Range | 100 MHz to 3 GHz |
| Resolution | 1 Hz |
| Number of Significant Digits | 9 digits |
| Accuracy | 1 microsecond acquisition window: +/- 1000 ppm 10 microsecond acquisition window: +/- 100 ppm 100 microsecond acquisition window: +/- 10 ppm > 1 millisecond acquisition window: +/- 2.5 ppm Note: min. acquisition window: 150 ns |
| Power Measurement | |
| Range | -34 dBm to + 16 dBm |
| Resolution | .01 dBm |
| Accuracy | +/- .25 dB, -20 dBm to +10 dBm +/- 1.00 dB, -34 dBm to +16 dBm Note: min. RF pulse width is 500 ns |
| Pulse Repetition Interval Measurement | |
| Range | 1 microsecond to 100 milliseconds |
| Resolution | 1 nanosecond |
| Accuracy | ± 1 nanosecond |
| Pulse Width | |
| Range | 100 nanoseconds to 10 milliseconds |
| Resolution | 1 nanosecond |
| Accuracy | ± 1 nanosecond |
| Time Base | |
| Standard | 10 MHz TCXO or PXIbus CLK10 |
| Accuracy | ±2 ppm |
| Power Requirements (Typical) | +5 V @ 1.85 Amp +3.3 V @ .2 Amp |
| Operating Temperature Range | -20° C to +60° C |
| Storage Temperature Range | -40° C to +75° C |
| Size | 3U PXI |
| Weight | 16 oz. |
Reseñas