| | BT3563/BT3563-01 | BT3562/BT3562-01 |
| Max. applied measurement voltage | }300 V DC rated input voltage }300 V DC maximum rated voltage above ground | }60 V DC rated input voltage }70 V DC maximum rated voltage above ground |
| Resistance Measurement ranges | 3 mƒ¶ (Max. display 3.1000 mƒ¶, resolution 0.1 ƒÊƒ¶) to 3000 ƒ¶ (Max. display 3100.0 ƒ¶, resolution 100 mƒ¶), 7 ranges Accuracy: 30 mƒ¶ to 3000ƒ¶ ranges, }0.5 % rdg. }5 dgt. (Add }3 dgt. for EX.FAST, or }2 dgt. for FAST and MEDIUM) 3 mƒ¶ range, }0.5 % rdg. }10 dgt. (Add }30 dgt. for EX.FAST, or }10 dgt. for FAST, or }5 dgt. for MEDIUM) Testing source frequency: 1kHz }0.2Hz, testing current: 100 mA (3 mƒ¶ range) to 10 ƒÊA (3000ƒ¶ range) Open-circut Voltage: 25 V peak (3/30 mƒ¶ ranges), 7 V peak (300 mƒ¶ range), 4 V peak (3ƒ¶ to 3000ƒ¶ ranges) |
| Voltage Measurement ranges | DC 6 V (resolution 10 ƒÊV) to DC 300 V (resolution 1 mV), 3 ranges | DC 6 V (resolution 10 ƒÊV) to DC 60 V (resolution 100 ƒÊV), 2 ranges |
| Accuracy: }0.01 % rdg. }3 dgt. (Add }3 dgt. for EX.FAST, or }2 dgt. for FAST and MEDIUM) |
| Display | 31000 full digits (resistance), 600000 full digits (voltage), LED |
| Sampling rate | Four steps 4 ms(Extra-FAST), 12 ms(FAST), 35 ms(Medium), 150 ms(Slow) (Typ., sampling time depends on supply frequency settings and function.) |
| Measurement time | Response time + sampling rate, approx. 10 ms for measurements (Response time depends on reference values and the measurement object.) |
| Comparator functions | Setting: Upper and lower limit Output: 3 levels (Hi, In, Lo) or (Pass, Fail), Open-collector(35V, DC50mA max), Display, Dual audible indicator |
| Analog Output | Mesured resistance (displayed value, from 0 to 3.1V DC, -01 suffix models only) |
| Interface | External I/O, RS-232C, Printer (RS-232C), GP-IB(-01 suffix models only) |
| Power supply | 100 to 240 V AC, 50/60 Hz, 30 VA max. |
| Dimensions, mass | 215 mm (8.46 in) W ~ 80 mm (3.15 in) H ~ 295 mm (11.61 in) D, 2.4 kg (84.7 oz) |