Improve stability and reduce calibration costs compared with the previous HIOKI model
For Instrumentation systems (4 - 20 mA) and loop testing
Check temperature control equipment and electric distribution
8 types of thermocouples to test thermoelectric power generation
Ideal for electrical device testing and routine maintenance of production equipment such as calibrators
Use the max. 25 mA DC sink as an electric load
Circuit method
Bipolar sink and source
Constant Voltage
2.5 V: 0 to }2.5000 V (}0.03 % of setting }300 ƒÊV, 100 ƒÊV resolution) 25 V: 0 to }25.000 V (}0.03 % of setting }3 mV, 1 mV resolution)
Constant Current
25 mA: 0 to }25.000 mA (}0.03 % of setting }3 ƒÊA, 1 ƒÊA resolution)
Thermoelectric power generation
K: at TC: 0 ‹C, -174.0 to 1372.0 ‹C (}0.05 % of setting }0.5 ‹C, 0.1 ‹C resolution), Other types: E, J, T, R, S, B, N selectable
Thermoelectric power generation
K: at TC: RJ, -174.0 to 1372.0 ‹C (}0.05 % of setting }1.0 ‹C, 0.1 ‹C resolution), Other types: E, J, T, R, S, B, N selectable
Standard resistance (Rs)
100 Ħ (}0.2 Ħ)
Automatic generation
Number of memory steps: 20 Interval time: 1 to 99 sec (at CV, CC, TC mode)
[Measurement functions]
Voltage
2.5 V: 0 to }2.8000 V (}0.03 % rdg. }300 ƒÊV, 100 ƒÊV resolution, 1 Mƒ¶ input resistance) 25 V: 0 to }28.000 V (}0.03 % rdg. }3 mV, 1 mV resolution, 1 Mƒ¶ input resistance)
Current
25 mA: 0 to }28.000 mA (}0.03 % rdg. }3 ƒÊA, 1 ƒÊA resolution, 25 ƒ¶ input resistance)
Temperature
-25.0 to }80.0 ‹C (}0.5 ‹C at 23 }5 ‹C, 0.1 ‹C resolution, use with the RJ sensor 9184)
Sampling rate
Approx. 1.67 times/sec
Additional functions
Zero adjustment, Overflow display, USB communication, Monitor
Power supply
AC ADAPTER 9445-02/03 (100 to 240 V AC 50/60 Hz, 9 VA), Ni-MH battery HR6 ~ 4, 6 VA, (fully charged 2500 mAh Ni-MH batteries: 170 minutes continuous use), or LR6 (AA) alkaline battery ~ 4, 6 VA
Dimensions, mass
104 mm (4.09 in) W ~ 180 mm (7.09 in) H ~ 58 mm (2.28 in) D, 660 g (23.3 oz) (including LR6 ~ 4 batteries)