Sorensen XDL35-5

XDL35-5

Sorensen XDL35-5 DC fuente de alimentación de sobremesa controlada digitalmente 105 W, 0-35 V, 0-5 A, 3 ranges

Ametek
$
Codigo del Producto: XDL35-5
8 a 10 semanas

Caracteristicas Principales

  • Rangos de salida Multiples de voltaje y corriente
  • Entrada numerica directa por teclado o con el control rotatorio de voltaje y corriente
  • Terminales de sensado local o remoto
  • Teclas y leyendas del panel frontal iliminadas
  • Hasta 30 memorias de configuraciones de set-up
  • Indicadores de salida de potencia
  • Modo de proteccion para panel de control manual cuando se programa via interface
  • Proteccion contra sobre voltaje
  • Proteccion contra sobre corriente
  • Proteccion contra sobrecalentamiento
  • Terminales Sense protegidas

La serie Sorensen XDL representan la "próxima generación" de fuentes de poder de laboratorio de lato desempeño. Las XDL proveen multiples rangos para incrementar la salida de corriente cuando se usa menos votaje. Estas fuentes usan tecnologia lineal pura. En contraste con otras unidades controladas digitalmente, la serie XDL de Sorensen tiene control numerico y control rotatorio (ambos). La pantalla y el teclado se iluminan con leyendas que proveen informacion al instante de la configuracion y el estatus de la fuente.

Para mayor conveniencia esta serie tiene la capacidad de guardar en memoria no volatil hasta 10 configuraciones (30 para las de salida triple). Se tienen controles de sobre-voltaje y sobre-corriente.

La serie XDL tiene también capacidad de sensado remoto de voltaje, con terminales dedicadas para sensar la salida de manera remota.

El modelo XDL de salida triple ofrece funcionalidades para enlazar las dos salidas de la fuente de manera simultanea. Cuando estan enlazadas el teclado y la perilla rotatoria opera ambas salidas de manera simultanea para las salidas 1 y 2.

Electrical Specifications

Output ranges:

Range 1

0-35 V, 0-3 A

Range 2

0-15 V, 0-5 A

Range 3

0-35 V, 0-500.0 mA

Outputs

1

Output power

105 W

Interface (GPIB/RS-232/USB)

No

Voltage setting

By floating point numeric entry or rotary jog wheel; resolution 1 mV

Current setting

By floating point numeric entry or rotary jog wheel; resolution 1 mV or 0.1 mA depending on range

Setting accuracy

Voltage - 0.03% +/- 5 mV. Current 0.2% +/- 5 mA, 0.5mA

Output mode

Operation in constant voltage or constant current modes with automatice cross-over and mode indication by LEDs

DC output switch

Sets output voltage and current levels to zero when Off

Output terminals

4mm terminals on 19mm (0.75") spacing

Load regulation

Voltage: <0.01% + 2mV
Current: <0.01% + 250mA; <0.01% + 50mA on 500mA range (measured at output terminals using remote sense)

Line regulation

Voltage: <0.01% + 2mV for 10% line change
Current: <0.01% + 250mA; <0.01% + 50mA on 500mA range

Ripple and Noise

Typically <0.35mV rms 2m V p-p CV mode, and <0.2mA rms, <20mA rms (500 mA range) CI mode

Transient response

50ms to within 15 mV of set level for a change in load current from full load to half load or vice versa

Temperature coefficient

< +/- (50ppm + 0.5 mV)/oC (voltage)

Remote sense

Eliminates up to 0.5 V drop per lead. Remote sense operation selected from front panel and indicated by LED

Sense terminals

Recessed sprung sockets for direct insertion of wires. Duplicated on rear terminal block (P versions only)

General Specifications (1)

Power requirements

115 V or 230 V +/- 10% (adjustable internally, option HV for factory set 230 VAC input), 50/60 Hz. Installation Category ll

Operating range

5oC to 40oC, 20% to 80% RH

Storage range

-40oC to 70oC

Dimensions (HxWxD)

6.3 x 5.5 x 11.4"/160 x 140 x 290 mm

Weight

11.9 lbs/5.4 kg

Benchtop Operation

Folding legs are incorporated that can be used to angle the front panel upwards when required

Rack mount operation

19 inch 4U mount for up to three single output units or one triple plus one single; Blanking plates available for un-used sections

Warranty

3 years

Output Protection and Metering

Output protection

Output will withstand forward voltages of up to 20 V above rated output voltage. Reverse protection by diode clamp for current up to 3 A

Fault condition trip

The output will be shut down if any of the four trip conditions listed below occur. In addition to the output eing set Off, an isolated rear panel signal is also activated.

Over voltage (OVP)

Settable 1V to 40 V in 0.1 V steps

Over current (OCP)

Settable 0.1 A to 5.5 A in 0.01 A steps

Over temperature

Monitors internal temperature rise to protect against excess ambient temperature or blocked ventilation slots

Sense Error

Monitors the voltage between the remote sense terminals and output terminals to protect against mis-wiring

Trip output signal

Isolated open-collector output signal on rear panel

Meter resolution and accuracy:

Voltage (CI mode):

Resolution 10 mV Accuracy +/- (0.1% of reading + 10 mV)

Current (CV mode):

Resolution 0.001 A; 0.1 A on 500 mA range Accuracy +/- (0.2% + 0.005 A); +/- (0.2% + 0.5 mA) on 500 mA range

V x A

Resolution 0.01 W; 0.001 W on 500 mA range Accuracy +/- (0.3% + 0.05 W); +/- (0.3% +0 0.005 W) on 500 mA range

Store/Recall Settings

Number of stores

10 plus power-down store

Memory type

Non-volatile using EEPROM

Parameters stored

Range, Set volts, Set current, OVP, OCP

Recall system

Settings are previewed onto the displays before being actioned

Bus Interfaces (P Suffix versions)

USB

Standard USB hardware connection Supplied with device driver for Win 98 or above. Operates as a virtual COM port.

RS-232

Variable baud rate 19,200 max. Single instrument or Addressable RS-232

GPIB

Conforming with IEEE-488.1 and IEEE-488.2 (N.B. All three interfaces incorporate full control, readback and status reporting)

BUS type selection

From front panel (GPIB/RS-232/USB)

Address selection

From front panel (1 to 31)

Baud selection

RS-232 only. Fron front panel (600 to 19200 baud)

Setting resolution

Voltage - 1mV, Current 0.1 mA (0.1 mA on 500 mA range)

Accuracy

See specifications under Outputs and Metering

Remote control response time:

Interface

Typically < 80 ms

Output voltage

Response time varies with range and load conditions. Typical time to settle to within 1% of the total excursion on a 35 V / 3 A range with full load is <25 ms. With no load it is <7 ms for an upward charge and < 600 ms for downward.

Regulatory Approvals

CE-marked units meet: EN61010-1 and EN61326

1. General specifications apply for 5 to 4°Ctemperature range. Accuracy specifications apply for 18 to 28°C range temperature range after 1 hour warm-up with no load and calibration at 23°C.

Typical specifications are determined by design and are not guaranteed.

Note: Specifications are subject to change without notice.